Setting the target standard deviation for proficiency testing assessment by Algorithm S

Wannee Aupaiboon, Rachada Hemapattawee, Pochaman Tagheen

Abstract


This study aims to determine the target standard deviation for proficiency assessment of total dissolved solids in water scheme. Standard deviations from previous proficiency schemes which conducted during 2008 to 2014 at the concentration range of 300 - 400 mg/L were collected. The target standard deviations for proficiency assessment were calculated using Algorithm S (ISO 13528:2005) approach. It is found that the target standard deviation is 5.0 %RSD. This is corresponding with the technical criteria for evaluate laboratories’ performance. Acceptable target standard deviation in the same range of the sample concentration was achieved. This approach can be applied to the other concentration range and other sample matrix for proficiency testing scheme.

 

บทคัดย่อ


การศึกษานี้เป็นการหาส่วนเบี่ยงเบนมาตรฐานเป้าหมายของกิจกรรมทดสอบความชำนาญรายการสารที่ละลายได้ทั้งหมดในน้ำ โดยใช้ข้อมูลค่าส่วนเบี่ยงเบนมาตรฐานโรบัสต์จากการจัดกิจกรรมการทดสอบความชำนาญที่จัดขึ้นในระหว่างปี พ.ศ. 2551 ถึง พ.ศ. 2557 ที่มีค่าความเข้มข้นของตัวอย่างอยู่ในช่วง 300 - 400 mg/L นำมาหาค่าของส่วนเบี่ยงเบนมาตรฐานเป้าหมายของกิจกรรมโดยใช้รูปแบบสถิติ Algorithm S (ISO 13528:2005) มาประยุกต์ใช้ในการคำนวณ พบว่าค่าส่วนเบี่ยงเบนมาตรฐานเป้าหมายที่ได้มีค่าเท่ากับ 5.0 %RSD ซึ่งสอดคล้องกับเกณฑ์ทางเทคนิคในการประเมินสมรรถนะห้องปฏิบัติการ ผลจากการวิจัยครั้งนี้ทำให้ได้ค่าส่วนเบี่ยงเบนมาตรฐานเป้าหมายที่เหมาะสมสำหรับช่วงความเข้มข้นของตัวอย่าง ซึ่งสามารถนำแนวทางนี้ไปประยุกต์ใช้กับช่วงความเข้มข้นอื่นและตัวอย่างชนิดอื่นต่อไป